Tender

Procurement of a 350-1000 nm Tunable Pulsed Laser Source

  • National Physical Laboratory

F02: Contract notice

Notice identifier: 2023/S 000-034239

Procurement identifier (OCID): ocds-h6vhtk-041b35

Published 20 November 2023, 12:00pm



Section one: Contracting authority

one.1) Name and addresses

National Physical Laboratory

Hampton Road

Teddington

TW11 0LW

Contact

Nina Heath

Email

nina.heath@npl.co.uk

Country

United Kingdom

Region code

UK - United Kingdom

Internet address(es)

Main address

www.npl.co.uk

one.3) Communication

The procurement documents are available for unrestricted and full direct access, free of charge, at

https://lupc.bravosolution.co.uk/

Additional information can be obtained from the above-mentioned address

Tenders or requests to participate must be submitted electronically via

https://lupc.bravosolution.co.uk/

Tenders or requests to participate must be submitted to the above-mentioned address

one.4) Type of the contracting authority

Body governed by public law

one.5) Main activity

Other activity

Research


Section two: Object

two.1) Scope of the procurement

two.1.1) Title

Procurement of a 350-1000 nm Tunable Pulsed Laser Source

two.1.2) Main CPV code

  • 38000000 - Laboratory, optical and precision equipments (excl. glasses)

two.1.3) Type of contract

Supplies

two.1.4) Short description

The National Physical Laboratory requirement is for the supply of a laser source continuously tunable over the 350-1000 nm wavelength range. The source is to be used to calibrate single-photon detectors (SPDs) in NPL’s new traceable calibration facility.

The source is required to be pulsed at a variable repetition rate.

The pulse jitter must be short enough to allow timing jitter measurements of SPDs.

The pulse width must be short enough to enable measurement of the detection profiles of gated detectors.

The pulse-to-pulse energy variability must be low enough so that the intensity of individual pulses do not need to be measured, and the mean value represents a very good approximation to the energy of any individual pulse.

The technologies likely to fulfil these requirements are Optical Parametric Oscillators (OPOs) with a Second Harmonic Generator (SHG) module and a stable pump laser to allow a low pulse-to-pulse amplitude uncertainty, or an Optical Parametric Amplifier (OPA) with similar SHG and stable pump laser requirement.

The 350-1000 nm wavelength tuning range is required to be achieved without change of optics or re-alignment. A pulse repetition rate between 1 MHz and 100 kHz is required (10 MHz-100 kHz is desirable). The operational temperature range is at least from 15 C to 30 C.

two.1.5) Estimated total value

Value excluding VAT: £270,000

two.1.6) Information about lots

This contract is divided into lots: No

two.2) Description

two.2.3) Place of performance

NUTS codes
  • UK - United Kingdom
Main site or place of performance

Teddington, UK

two.2.4) Description of the procurement

The National Physical Laboratory requirement is for the supply of a laser source continuously tunable over the 350-1000 nm wavelength range. The source is to be used to calibrate single-photon detectors (SPDs) in NPL’s new traceable calibration facility.

The source is required to be pulsed at a variable repetition rate.

The pulse jitter must be short enough to allow timing jitter measurements of SPDs.

The pulse width must be short enough to enable measurement of the detection profiles of gated detectors.

The pulse-to-pulse energy variability must be low enough so that the intensity of individual pulses do not need to be measured, and the mean value represents a very good approximation to the energy of any individual pulse.

The technologies likely to fulfil these requirements are Optical Parametric Oscillators (OPOs) with a Second Harmonic Generator (SHG) module and a stable pump laser to allow a low pulse-to-pulse amplitude uncertainty, or an Optical Parametric Amplifier (OPA) with similar SHG and stable pump laser requirement.

The 350-1000 nm wavelength tuning range is required to be achieved without change of optics or re-alignment. A pulse repetition rate between 1 MHz and 100 kHz is required (10 MHz-100 kHz is desirable). The operational temperature range is at least from 15 C to 30 C.

two.2.5) Award criteria

Price is not the only award criterion and all criteria are stated only in the procurement documents

two.2.6) Estimated value

Value excluding VAT: £270,000

two.2.7) Duration of the contract, framework agreement or dynamic purchasing system

End date

30 August 2024

This contract is subject to renewal

No

two.2.10) Information about variants

Variants will be accepted: Yes

two.2.11) Information about options

Options: No

two.2.13) Information about European Union Funds

The procurement is related to a project and/or programme financed by European Union funds: No


Section four. Procedure

four.1) Description

four.1.1) Type of procedure

Open procedure

four.1.8) Information about the Government Procurement Agreement (GPA)

The procurement is covered by the Government Procurement Agreement: Yes

four.2) Administrative information

four.2.2) Time limit for receipt of tenders or requests to participate

Date

5 January 2024

Local time

5:00pm

four.2.4) Languages in which tenders or requests to participate may be submitted

English

four.2.7) Conditions for opening of tenders

Date

8 January 2024

Local time

9:00am


Section six. Complementary information

six.1) Information about recurrence

This is a recurrent procurement: No

six.4) Procedures for review

six.4.1) Review body

NPL Management

Teddington

Country

United Kingdom